Indirect Estimation of Total Prostatic Specific Antigen (TPSA) on Glassy Carbon Electrode Through its Interaction with Zinc Acetate Using Differential Pulse Voltammetric Technique
DOI:
https://doi.org/10.5281/zenodo.16644190Abstract
Abstract Views: 279
An indirect electrochemical behaviour of total prostatic specific antigen (TPSA) was studied using the differential pulse technique coupled with a three-electrode detection system, which consists of 2mm diameter glassy carbon as working electrode, 2mm diameter platinum wire as an auxiliary electrode and silver/silver chloride saturated potassium chloride as reference electrode (Ag/AgCl sat KCl). The method is based on monitoring the decrease in the zinc acetate reduction peak current, which is due to its interaction with TPSA. Acetate buffer solution (pH4) was used, the potential scanned between -1.4 V to -0.5V, and the optimum conditions were tested. The calibration curve for TPSA was constructed for a sequence addition of TPSA, and it was linear within the studied concentration range (2.4431- 24.4554) x10-9 M, with a correlation coefficient of 0.9763 and calibration equation y= 0.0308x + 1.287. The binding constant (k) and the thermodynamic parameters (ΔG, ΔH, and ΔS) for Zn2+- TPSA interaction were calculated.
Keywords:
Differential pulse voltammetry, Glassy carbon, Indirect determination, Total Prostatic Specific AntigenReferences
Al-Hasany, A. M., Al-Taee, A. T., & Al-Wahb, H. A. (2019). Electrochemical study of theophylline-creatinine interaction using square wave voltammetry. The Eurasia Proceedings of Science Technology Engineering and Mathematics, 7, 204-211.
Al-Hyali, R. H., & Al-Taee, A. T. (2021). A New GE/MWCNTs/PFA modified electrode for simultaneous determination of catechol and hydroquinone. Research Journal of Pharmacy and Technology, 14(2), 828-832.
Al-Imam, M. M. A., & Al-Abady, F. M. (2022, November). Electrochemical behavior of aspartic amino acid and its molecular interaction with Mn (II) ions at different conditions using SWV method. In AIP Conference Proceedings (Vol. 2394, No. 1, p. 040055). AIP Publishing LLC. https://doi.org/10.1063/5.0121576 .
Al-Taee, A. T., & Al-Hafidh, A. Z. (2021). Indirect electrochemical determination of chlorpropamide through its interaction with valsartan using square wave voltammetry. Research Journal of Pharmacy and Technology, 14(12), 6541-6544. http://dx.doi.org/10.52711/0974-360X.2021.01131
Bray, F., Ferlay, J., Soerjomataram, I., Siegel, R. L., Torre, L. A., & Jemal, A. (2018). Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA: A Cancer Journal for Clinicians, 68(6), 394-424. https://doi.org/10.3322/caac.21492
Damborska, D., Bertok, T., Dosekova, E., Holazova, A., Lorencova, L., Kasak, P., & Tkac, J. (2017). Nanomaterial-based biosensors for detection of prostate specific antigen. Microchimica acta, 184(9), 3049-3067. https://doi.org/10.1007/s00604-017-2410-1
Filik, H., Avan, A. A., & Özyürek, M. (2021). Electrochemical immunosensors based on nanostructured materials for sensing of prostate-specific antigen: a Review. Current Medicinal Chemistry, 28(20), 4023-4048. https://doi.org/10.2174/0929867328666201124151821
Hasan, M. Q., Al-Taee, A. T., & Al-Imam, M. M. (2024). Electrochemical Behavior of Finasteride on Glassy Carbon Electrode Using Differential Pulse Voltammetric Technique. Analytical and Bioanalytical Electrochemistry, 16(8), 752-763.
Healy, D. A., Hayes, C. J., Leonard, P., McKenna, L., & O’Kennedy, R. (2007). Biosensor developments: application to prostate-specific antigen detection. TRENDS in Biotechnology, 25(3), 125-131. https://doi.org/10.1016/j.tibtech.2007.01.004
Li, S., Zhang, J., Tan, C. S., Chen, C., Hu, C., Bai, Y., & Ming, D. (2022). Electrochemical immunosensor based on hybrid MoS2/Pt@ Au-nanoprism/PDA for simultaneous detection of free and total prostate specific antigen in serum. Sensors and Actuators B: Chemical, 357, 131413. https://doi.org/10.1016/j.snb.2022.131413
Liu, S., Zhang, X., Wu, Y., Tu, Y., & He, L. (2008). Prostate-specific antigen detection by using a reusable amperometric immunosensor based on reversible binding and leasing of HRP-anti-PSA from phenylboronic acid modified electrode. Clinica Chimica Acta, 395(1-2), 51-56. https://doi.org/10.1016/j.cca.2008.04.031
Ludwig, J. A., & Weinstein, J. N. (2005). Biomarkers in cancer staging, prognosis and treatment selection. Nature Reviews Cancer, 5(11), 845-856. https://doi.org/10.1038/nrc1739
Mascini, M., & Tombelli, S. (2008). Biosensors for biomarkers in medical diagnostics. Biomarkers, 13(7-8), 637-657. https://doi.org/10.1080/13547500802645905
Mitrakas, M., & Ochsenkühn-Petropoulou, M. (2021). The 12th International Conference on instrumental analysis—modern trends and applications (20–23 September 2021, virtual event). Applied Sciences, 11(24), 11767. https://doi.org/10.3390/app112411767
Morrin, A., Killard, A. J., & Smyth, M. R. (2003). Electrochemical characterization of commercial and home-made screen-printed carbon electrodes. Analytical letters, 36(9), 2021-2039. https://doi.org/10.1081/AL-120023627
Parnsubsakul, A., Safitri, R. E., Rijiravanich, P., & Surareungchai, W. (2017). Electrochemical assay of proteolytically active prostate specific antigen based on anodic stripping voltammetry of silver enhanced gold nanoparticle labels. Journal of Electroanalytical Chemistry, 785, 125-130. https://doi.org/10.1016/j.jelechem.2016.12.010
Prensner, J. R., Rubin, M. A., Wei, J. T., & Chinnaiyan, A. M. (2012). Beyond PSA: the next generation of prostate cancer biomarkers. Science Translational Medicine, 4(127), 127rv3-127rv3. https://doi.org/10.1126/scitranslmed.3003180
Thakur, A., & Kumar, A. (2023). Exploring the potential of ionic liquid-based electrochemical biosensors for real-time biomolecule monitoring in pharmaceutical applications: From lab to life. Results in Engineering, 20, 101533. https://doi.org/10.1016/j.rineng.2023.101533
Tsoi, T. H. (2017). Novel biosensing platforms for a fast screening of milk contaminants and prostate cancer (Doctoral Dissertation, Hong Kong Polytechnic University).
Vanova, V., Mitrevska, K., Milosavljevic, V., Hynek, D., Richtera, L., & Adam, V. (2021). Peptide-based electrochemical biosensors utilized for protein detection. Biosensors and Bioelectronics, 180, 113087. https://doi.org/10.1016/j.bios.2021.113087
Zheng, Y., Yuan, Y., Chai, Y., & Yuan, R. (2015). A label-free electrochemical aptasensor based on the catalysis of manganese porphyrins for detection of thrombin. Biosensors and Bioelectronics, 66, 585-589. https://doi.org/10.1016/j.bios.2014.12.022
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2025 Biomedicine and Chemical Sciences

This work is licensed under a Creative Commons Attribution 4.0 International License.